Data Analysis for Quantitative PCR

Home > DNA Extraction & Quantitation > Quantitation > Data Analysis for Quantitative PCR
Quantitative PCR Amplification Graph

Quantitative PCR monitors the increase in fluorescent signal throughout the PCR cycling process. The change in fluorescence is monitored between samples and standards so that a comparison can be made.12

The amplification process can be summarized in three consecutive phases:

  1. Exponential/ Geometric amplification
  2. Linear amplification
  3. Plateau region
Phase One: Exponential/Geometric

During the exponential phase of the PCR process, the reaction results in a theoretical doubling of amplicons with each cycle.  While the ideal efficiency is not achieved, the doubling is close to 100%, yielding a consistent relationship of input DNA to product.13

At the beginning of the exponential phase, the baseline is determined by measuring the background fluorescence signal (noise). This baseline establishes the threshold. During the amplification process of a sample, the point (in terms of amplification cycles) in which the level of fluorescence exceeds the threshold is referred to as the Cycle Threshold (CT) . The CT value is lower for a sample with a higher initial concentration and is higher for a lower concentration sample.

standard curve

The standard curve is a regression line that is derived from the CT values of the standards plotted against the log of the concentration of the standards. The concentration of each sample is determined by comparing its CT value against the standard curve.

Phase Two: Linear Phase

The consumption of one or more PCR reagents during the amplification process will impede the efficiency causing the onset of the linear phase. Product amplification slows down yielding an inconsistent ratio of input DNA to product. Therefore, the linear phase is not commonly used in the data analysis process.

Phase Three: Plateau Region

The final amplification phase signifies the depletion of critical reagents and is known as the plateau region. During this phase, the amplification process ceases.14