President's DNA Initiative
-
Advancing Justice Through DNA Technology
DNA Analyst Training
previous pagenext page

Primer Annealing Temperature

Home > DNA Amplification > Overview > Primer Design > Primer Annealing Temperature

The annealing temperature (Ta) chosen for a PCR depends directly on length and composition of the primer(s). Generally, an annealing temperature about 5°C below the lowest Tm of the pair of primers is used.11 The optimal annealing temperature for any given primer pair on a particular target can be calculated as follows:12

Ta Opt = 0.3 x(Tm of primer) + 0.7 x(Tm of product) - 25

  • Tm of primer is the melting temperature of the less stable primer-template pair.
  • Tm of product is the melting temperature of the PCR product.

Polymerases require the binding of nucleotides at the 3′ end of the primer to begin elongation, and because of this, any nonspecific binding at the 3′ end will adversely affect amplification. Non-specific binding that occurs at the 5′ end of the primer does not necessarily adversely impact amplifications since polymerases cannot begin elongation until the 3′ end binds.

annealed primer

One consequence of having too low a Ta is that one or both primers will anneal to sequences other than the true target, as internal single-base mismatches or partial annealing may be tolerated. This can lead to nonspecific amplification and will consequently reduce the yield of the desired product if the 3′ -most base is paired with a target. Conversely, too high a Ta may yield little product, as the likelihood of primer annealing is reduced.

Use of this calculated optimal Ta in the annealing step of the PCR cycle usually results in optimal PCR product yield with minimum false product production.

< Previous Page  ::  Next Page >

© 2007 NFSTC Science Serving Justice®
NOTE TO USERS: The President’s DNA Initiative DNA Analyst Training program and assessment were completed and published in 2005, in cooperation with the National Institute of Justice. The science and techniques in the program are sound and proven, however, program content has not been updated to include tools and technologies developed and in use after 2005, including many kits and robots. Assessment questions address only content delivered in this program and may not contain the full range of tools in use in your laboratory.